Quick Answer: For tablet hardness testing during an A/B rotary press trial, first verify sample identity, tablet shape, loading direction, and the test method. Then compare individual breaking-force readings with tablet weight, thickness, and the press records for the same collection interval. Apply the product’s agreed acceptance criteria before deciding whether a press adjustment is justified.
Two sample trays can come from the same batch and still give an unfair comparison. One may contain startup tablets; the other may have been tested with a different loading setting. This guide takes the trial from preparation and sampling through result review, with published research data and a checklist for your own A/B records.
What Does Tablet Hardness Testing Measure?
In this article, tablet hardness means the force required to fracture a tablet in a specified plane. The tester reports a breaking-force result, commonly in newtons (N). [4] In the trial record, attach that result to the tablet’s identity and the method used to break it.
A hardness result answers one question about the tablet. Keep the following measurements separate when deciding what the trial has demonstrated.
| Measurement | What it describes | How to use it in the trial |
|---|---|---|
| Breaking force, N | The force that fractures an individual tablet. | Retain each result, placement, and fracture observation. |
| Friability, % mass loss | Sample mass lost during a defined tumbling test. [2] | Review separately when tablets chip or shed material. |
| Disintegration, time | How long a tablet takes to break into smaller particles. [4] | Include the product’s required test in the trial review. |
| Dissolution | Drug release into solution under the specified test conditions. [4] | Keep the applicable release results with the other quality records. |
If transfer damage is the concern, review tablet friability testing alongside the breaking-force data and observations at the transfer points. A single N value does not report what happened during those separate operations.
The acceptance criterion belongs to the product. USP 〈1217〉 explains that acceptable strength depends on intended use and the release mechanism. [1] Bring the relevant specification to the trial. If a development target is still being investigated, identify it as provisional in the report.
What Should You Prepare Before an A/B Press Trial?
Write a one-sentence trial question before collecting samples: which planned change are A and B intended to compare? For a compression-force comparison, name that change explicitly and identify the conditions that should remain constant. If the plan changes force and speed together, record it as a combined change.
Agree on these items with the production and quality team before the run:
- Material and tooling identity. Record the product, formulation revision, batch, press, and punch/die identification. Note the tablet shape, nominal dimensions, face profile, and score line.
- The A/B settings. List the planned difference and the other settings to retain, including precompression, turret speed, and feed settings where applicable. Keep room for actual recorded values beside the plan.
- The sampling plan. Specify the collection point, sample count, intervals, and run stages. Define when the run is ready for the planned comparison; do not decide this retrospectively after seeing favorable readings.
- The measurement method. Identify the tester, method revision, applicable calibration or verification record, loading mode, rate with units, and placement instruction.
- The review rule. Identify the acceptance criteria and required companion tests. Decide how to document missing readings, unusual breaks, and additional sampling.

Loading settings deserve an explicit check. SOTAX’s technical note explains that different force-application settings and tablet alignment can change breaking-force results. A constant-speed setting and a force-increase-rate setting describe different controls; record the selected mode and its actual value. [5]
Also plan sample handling. Label the containers before collection, record collection and test times, and follow the holding conditions in the product method. An elapsed-time difference should remain visible in the A/B record rather than being lost when the results are averaged.
How to Run a Tablet Hardness Test During a Press Trial
Use the same recording sequence for both conditions. Keep startup, adjustment, and the protocol-defined stable-run samples in separate sets. The following workflow organizes the trial; the product method supplies the sample count and test settings.
- Identify the collection interval. Record the time, sampling point, stage, and press-data reference. Assign a group label and individual IDs such as A-01 and A-02. Repeat the same scheme for B.
- Inspect the tablets before testing. Note visible damage, irregular edges, or an existing crack. Preserve that observation so it is distinguishable from damage produced by the test.
- Measure before the destructive test. Record each tablet’s weight and thickness under its own ID. Add diameter when the agreed analysis uses it. Keep the measurements together as the tablet moves between instruments.
- Place the tablet using the documented reference. Describe the loading direction against an identifiable tablet feature. For a scored tablet, specify the score’s orientation relative to the load; a placement photograph or sketch can remove ambiguity.
- Run the selected method. Follow the tester’s operating instructions and record the result and test time. Check the method settings again if instruments or operators change during the trial.
- Save the original output. Link the instrument record, any fracture image, and the corresponding press-data interval to each tablet’s record. Preserve unsuccessful tests as documented events.
How to Record Tablet Breaking Force in Newtons
Use one measurement unit consistently. Keep the tester’s original reading and identify any conversion in the calculation record. Do not copy a number from an older report without checking the unit beside it.
Write the fracture observation next to the N value: a split across the diameter, separation along a score, edge chipping, or another observed break. Avoid replacing the observation with a conclusion such as “bad compression” before the records have been reviewed.
If the tester reports an error or produces no valid break reading, enter the error and the reason the value is missing. Do not substitute zero. If additional tablets are tested, give them new IDs and retain the link to the original event.
How to Compare Tablet Hardness Across Press Settings
Review the comparison in three passes. First verify that the intended samples and methods were used. Next review the individual measurements and press records together. Then apply the acceptance rule. This sequence prevents a difference in the testing setup from being mistaken for evidence that a machine adjustment worked.
What Changes When Tablet Compression Force Changes?
Main compression force (kN) describes the forming load recorded at the press. Tablet breaking force (N) comes from the subsequent destructive tablet test. Keep them in separate columns, as the published rotary-press study does. [2] Converting the units does not convert one measurement into the other.
Obtain the press value from its force-monitoring display or trial export where available. Label it as an actual measurement, an interval summary, or a target setpoint. If the press does not provide measured force, state that limitation; an adjustment position or maximum machine rating is not a substitute.
Match each sample set to its collection interval. Where several tablets share an interval-average force, label that shared basis clearly. Do not imply that each tablet has an individually traced forming-force value unless the data actually support that traceability.
Check weight, thickness, speed, and other recorded conditions before attributing a result to main compression. If several variables moved, describe the combined change and identify what a controlled follow-up would need to hold constant.
Review Individual Results Before the Average
Keep the actual number tested, individual readings, group mean, and spread together. A mean describes the center of a sample set; it cannot identify which tablet was unusually weak or whether several specimens fractured differently.
Review the values in collection order as well as by group. A cluster of unusual readings near an adjustment calls for a different follow-up from an isolated reading with a documented placement problem. State the observation first, then identify the record or additional check needed to investigate it.
Apply the product’s stated rule to the relevant results. If a rule concerns individual tablets, an acceptable mean cannot establish that every individual result met it. If the protocol calls for repeat intervals or further tests, complete those steps before writing that the setting is suitable.
When Does Tablet Tensile Strength Help?
For suitable flat-faced round tablets loaded across the diameter, FDA’s chewable-tablet guidance gives the diametral tensile-strength relationship: [4]
Here, F is breaking force in N, D is diameter in mm, and t is thickness in mm. The result is N/mm², or MPa. Loading acts across the diameter, not through the thickness. The unmodified equation does not cover convex or irregular tablets. [4] Scored tablets also require care because the fracture mechanism may differ. [3]

The equation explains why dimensions matter: with F and D unchanged, a greater thickness gives a lower calculated value. If the method uses this calculation, enter the dimensions measured on each identified tablet before it was broken. Do not combine one tablet’s force with another tablet’s dimensions.
What Do Published Tablet Studies Show?
The following research examples illustrate two different review problems: interpreting a change in press conditions and checking how the tablet was tested.
Case 1: Compression Force at a Fixed Turret Speed
Mohylyuk and Bandere (2023) made 6 mm flat-faced isoniazid tablets on a Korsch XL 100 at 80 rpm, with precompression of 0.40 ± 0.03 kN. An Erweka TBH 20 measured breaking force within 12 hours after compaction. [2]
Across the eight reported conditions, mean breaking force rose from 60 N at 4.8 kN to 91 N at 14.0 kN; the 15.5 kN group averaged 87 N. This last comparison alone does not establish a statistically significant decline or an optimum setting. [2]

Compare the spread and dimensions alongside force. The complete published group summaries appear in the trial-report section below; your own A/B report also needs the underlying individual tablet records.
Case 2: Score Orientation and Fracture Behavior
Podczeck, Newton, and Fromme (2014) combined modeling with physical tests of commercially manufactured, flat, round, bevel-edged scored tablets. They examined five score orientations. The physical tests did not reproduce the large breaking-load differences predicted by the model, but they did show differences in fracture behavior. [3]
This was a laboratory investigation of testing geometry. Its lesson for a press trial is to preserve orientation and fracture observations even when force readings appear similar. Before interpreting an A/B difference, check that placement followed the same instruction. The study does not support a rule that turning every scored tablet will raise or lower its result by a predictable amount.
What Should You Check When Results Differ?
Use the table below to select the next check. These are investigation prompts, not diagnoses or automatic instructions to raise or lower the press setting.
| Observation | Check first | Useful next record or action |
|---|---|---|
| The group means differ | Sample stage, test settings, orientation, elapsed time, and dimensions. | Mark differences in the comparison record before assigning a cause. |
| The means are similar, but one group varies more | Individual readings, collection order, and fracture notes. | Identify where the variation occurs and plan the follow-up required by the protocol. |
| One tablet gives a low result or an unusual break | Its ID, pre-test condition, placement, dimensions, and instrument message. | Retain the result and investigate the event before summarizing the group. |
| Different testers give different readings | Method revision, loading mode, rate, alignment, and verification records. | Resolve method differences before using the sets as a press comparison. |
| Breaking force meets the criterion, but transfer damage remains | Friability results and observations at the actual transfer points. | Record when and where damage appears; review both tablet and handling evidence. |
For capping, lamination, or edge damage, describe what you can see and use tablet defects and their causes to organize the investigation. A photograph tied to a sample ID is more useful than a general note saying “tablets failed.”
If further testing is justified, document the reason, preserve the original output, and collect additional tablets under the quality procedure. A broken tablet cannot supply a second independent breaking-force result. Keep the new sample IDs and results linked to the first investigation rather than replacing the initial rows.
End the review with a specific next step: correct a documented test-method difference, repeat a defined collection interval, investigate a material or tooling observation, or evaluate a planned press change. The evidence should determine which step is appropriate.
What Belongs in the Trial Report?
Build the report in two parts: common conditions for the run and one row for each tested tablet. This keeps repeated information out of the measurement grid without losing the context needed to compare A and B.
Document the Common Trial Conditions
The right-hand column explains what to enter in your report; it contains recording guidance rather than measured results.
| Common field | What to record |
|---|---|
| Product and material | Product name, formulation revision, batch, trial date, protocol revision, and operator. |
| Press, tooling, and tablet | Press and tooling IDs; tablet shape, face profile, score, and nominal dimensions. |
| Conditions A and B | Planned change, retained settings, and actual press-data references, including precompression and feed settings where applicable. |
| Force-data basis | Data source and collection interval; identify measured values, interval summaries, and target setpoints. |
| Tester and method | Instrument ID, method revision, verification reference, loading mode and rate, and placement instruction. |
| Sampling and handling | Location, sample count, schedule, run-stage criteria, holding conditions, and planned sampling-to-test interval. |
| Acceptance and review | Product criteria and specification reference, companion tests, deviations, and additional-test records. |
Published Group Summaries: Eight Compaction Conditions
Each row below represents one condition in the published isoniazid study. Compaction force is the reported set condition; the measurements are group summaries.
| Compaction force (kN) | Breaking force, mean ± SD (N) | Mean weight (mg) | Mean thickness (mm) | Tensile strength, mean ± SD (MPa) |
|---|---|---|---|---|
| 4.8 | 60 ± 6.9 | 112.0 | 3.42 | 1.86 ± 0.21 |
| 6.0 | 65 ± 7.1 | 111.5 | 3.29 | 2.09 ± 0.23 |
| 8.0 | 73 ± 6.3 | 113.2 | 3.21 | 2.42 ± 0.21 |
| 9.0 | 78 ± 4.3 | 112.9 | 3.10 | 2.67 ± 0.15 |
| 9.9 | 83 ± 2.5 | 112.8 | 3.04 | 2.90 ± 0.09 |
| 11.3 | 89 ± 3.9 | 113.5 | 3.00 | 3.15 ± 0.14 |
| 14.0 | 91 ± 5.1 | 113.0 | 2.97 | 3.25 ± 0.18 |
| 15.5 | 87 ± 8.6 | 111.0 | 2.90 | 3.18 ± 0.32 |
Sample size and access: For the breaking-force column, n = 20 per condition. The paper directs data requests to the corresponding author. No per-tablet raw dataset was obtained for this article. [2]
For your own A/B trial, keep one row per tablet. Record its ID, collection stage and time, time to test, press-data interval and force-data basis, turret speed, weight, thickness, loading direction, breaking force, and fracture notes. Reference the common conditions when an interval value is shared across rows. Add measured diameter, calculated tensile strength, and the method reference when required.
After reviewing the individual records, summarize the actual sample count, mean, and spread for each comparable group. State which criteria were met, which observations remain unresolved, and what follow-up is required. Attach the original instrument output and press records so another reviewer can trace the conclusion.
If this work informs equipment selection, use the rotary tablet press selection guide to review tooling, tablet dimensions, and output requirements alongside the trial evidence.
Frequently Asked Questions
What is an acceptable tablet hardness value?
Use the criterion established for the product and test method. USP’s public introduction links acceptable strength to intended use and release mechanism. [1] A value copied from another product or a published research table does not establish your trial’s acceptance range.
How many tablets should be tested during a press trial?
Set the number and collection schedule in the product’s trial protocol before testing. The eight rows in the published table represent study conditions, not a sample count. Cover the required run stages and comparisons; one research paper’s sample size should not be adopted automatically.
Can A and B be tested on different hardness testers?
Check the method, loading mode and rate, alignment, and verification records first. SOTAX identifies loading conditions and positioning as important to comparability. [5] If equivalence has not been established, record the instrument change and resolve it before attributing a difference to the press.
Should the press force be increased after a low reading?
First trace the reading to its sample, dimensions, fracture notes, collection interval, and test method. Review the group and the product’s acceptance rule. A single result does not isolate the cause or establish that increasing main compression is the appropriate next step.
How soon after compression should tablets be tested?
Follow the interval and holding conditions in the product method and record the actual collection and test times. Flag departures in the report. The published study’s timing describes that experiment; it does not establish a universal waiting period for other formulations.
Can a damaged or already broken tablet be retested?
A tablet destroyed by the first test cannot provide a second independent breaking-force result. For damage observed before testing, follow the product method and document its condition. Where additional testing is justified, test newly identified specimens and retain the original records and reason for the follow-up.
To discuss rotary tablet press trial conditions with GRAND, send your tablet shape, target weight, existing product breaking-force criteria, and required output.
References
- United States Pharmacopeia. 〈1217〉 Tablet Breaking Force. Public introduction; product-dependent strength criteria.
- Mohylyuk V, Bandere D. High-Speed Tableting of High Drug-Loaded Tablets Prepared from Fluid-Bed Granulated Isoniazid. Pharmaceutics. 2023;15(4):1236. Sections 2.7, 2.9, 2.13; Table 3; Figure 7a; Data Availability Statement.
- Podczeck F, Newton JM, Fromme P. Theoretical investigations into the influence of the position of a breaking line on the tensile failure of flat, round, bevel-edged tablets …. International Journal of Pharmaceutics. 2014;477(1–2):306–316. DOI: 10.1016/j.ijpharm.2014.10.028. UCL repository record.
- U.S. Food and Drug Administration. Quality Attribute Considerations for Chewable Tablets: Guidance for Industry. August 2018. Sections III.A–C; Appendix I. Its recommendations address chewable tablets.
- SOTAX. Challenges in comparing breaking force test results. August 10, 2023. Technical note on loading conditions and tablet positioning.