Respuesta rápida: El Hausner ratio is tapped bulk density divided by untapped bulk density. The Carr index is 100 × (tapped density − untapped density) / tapped density. For densities of 0.500 and 0.625 g/mL, the results are 1.250 and 20.00%. Both describe relative powder consolidation and help screen materials before tablet pressing.
Puntos clave
- Hausner ratio (HR): Divide tapped bulk density by untapped bulk density.
- Carr index (CI): Express the decrease in bulk volume after tapping as a percentage of the initial volume.
- Relationship: Both indicators use the same paired measurement. They are mathematically related and do not provide two independent confirmations of flow.
- Interpretation: Lower values generally indicate less consolidation in comparable tests. No single HR or CI value proves suitable tablet press performance.
- Production: Retain the actual density values alongside the indices when planning die filling and tablet weight investigations.
What Is the Hausner Ratio?
The Hausner ratio compares the density of a powder bed after tapping with its original loose bulk density. It is dimensionless because both values use the same units. A ratio near 1 means that the measured bed volume changed relatively little; a larger value indicates more consolidation during the specified tapping procedure.
For pharmaceutical and nutraceutical teams, the ratio is an indirect indicator of material handling behavior. Particle geometry, fines, cohesion, moisture and the measurement procedure can all influence the result. USP General Chapter 〈1174〉 discusses the Hausner ratio together with the compressibility index as one approach to evaluating powder flow. [1]
What Is the Carr Index?
The Carr index, also called Carr’s compressibility index, expresses relative powder bed consolidation as a percentage. A 100 g sample that settles from 200 mL to 160 mL occupies 40 mL less apparent bulk volume, a reduction of 20%. In this illustrative calculation, the powder mass remains unchanged.
Compressibility in this context refers to the change observed during tapping. It does not measure the force needed to compact a blend into a tablet, the bonding strength of particles or the breaking force of the finished tablet. Avoid treating a flow index as a direct measure of tabletability.
Hausner Ratio Formula and Carr Index Formula
Use values measured on the same material portion. Let ρb denote untapped bulk density, ρt tapped bulk density, V0 untapped volume and Vf final tapped volume.
HR = V0 / Vf
CI = 100 × (V0 − Vf) / V0
These equivalent forms follow from density = mass / volume when the same sample retains its mass. [1] A frequent calculation error is dividing the density difference by untapped density instead of tapped density. That calculation produces a different percentage.
How the two indices are related
HR = 1 / (1 − CI / 100)
Keep full calculation precision until reporting the final values. Two results derived from the same unrounded density data must satisfy this relationship, subject to rounding. A discrepancy may reveal an input, unit or formula error. Neither index is an additional independent measurement.
How to Calculate Hausner Ratio and Carr Index
Continue the illustrative 100 g powder example from our Tap Density guide. The same sample occupies 200 mL before tapping and 160 mL at the documented final reading. The figures below are mathematical examples, not measurements from a GRAND customer or equipment trial.
| Input or result | Calculation | Valor |
|---|---|---|
| Sample mass | Same portion before and after tapping | 100 g |
| Untapped bulk density | 100 / 200 | 0.500 g/mL |
| Tapped bulk density | 100 / 160 | 0.625 g/mL |
| Hausner ratio | 0.625 / 0.500 | 1.250 |
| Carr index | 100 × (0.625 − 0.500) / 0.625 | 20.00% |
The volume equations produce exactly the same values: HR = 200 / 160 = 1.25 and CI = 100 × (200 − 160) / 200 = 20%. Under the referenced relative ranking these example results sit at the upper end of the fair-flow category.
Hausner Ratio and Carr Index Calculator
Choose one input method. Use paired values from the same powder sample in the same units. Results update automatically as you type.
Use a decimal point, with up to 9 digits before it and 6 digits after it. Hausner ratio is shown to 3 decimal places and Carr index to 2 decimal places.
Calculator loading. If results do not update, reload the page.
Calculated values are screening indicators. Interpret them using the laboratory’s documented method and reporting rules, then verify actual feeding and tablet quality.
Hausner Ratio and Carr Index Flowability Chart
USP 〈1174〉 includes a widely used descriptive ranking for the two indicators. Use the table to communicate relative flow characteristics, not as an automatic batch-release specification. [1]
| Carr index (%) | Hausner ratio | Relative flow |
|---|---|---|
| 1–10 | 1.00–1.11 | Excelente |
| 11–15 | 1.12–1.18 | Bien |
| 16–20 | 1.19–1.25 | Fair |
| 21–25 | 1.26–1.34 | Passable |
| 26–31 | 1.35–1.45 | Pobre |
| 32–37 | 1.46–1.59 | Very poor |
| Above 38 | Above 1.60 | Very, very poor |
The table follows the USP chapter effective May 1, 2024, including gaps at certain boundaries. A zero CI is possible if the measured volume is unchanged; report it as calculated rather than extending the table. Apply your laboratory’s documented reporting and rounding rules. If rounded values fall into different categories, retain both numerical results and explain the discrepancy.
Compare illustrative powder readings
| Bulk density (g/mL) | Tapped density (g/mL) | HR | CI |
|---|---|---|---|
| 0.500 | 0.550 | 1.100 | 9.09% |
| 0.500 | 0.625 | 1.250 | 20.00% |
| 0.400 | 0.600 | 1.500 | 33.33% |
A higher index can help prioritize an investigation into cohesion or packing variability, but the actual flow through a hopper or feed frame must be evaluated separately.
How to Obtain Comparable Test Results
Start with the laboratory’s established bulk and tapped density method. Record the initial and final volume of a representative powder portion, or record the corresponding densities calculated from that portion. Avoid mixing an untapped reading from one sample with a tapped reading from another.
| Control point | Por qué es importante |
|---|---|
| Sample identity and location | Connects the result to the same formulation and production stage |
| Sample handling | Transfer, vibration or holding time can change the initial packing state |
| Cylinder diameter and cylinder/holder mass | Can affect how the powder bed consolidates |
| Tapping apparatus and endpoint | Drop height, tap count and endpoint can affect the final volume |
| Powder mass and condition | Supports valid paired calculations and repeatability |
| Replicate measurements | Helps distinguish an unusual result from routine test variation |
In a 2019 interlaboratory study, Akseli and colleagues found that differences between lots were generally within the variability of the density test. Tapped density and HR showed no significant intersite bias in that study, while bulk density showed marginally significant bias. Record the method and individual readings before interpreting small differences between lots. [2] For the complete measurement sequence, see the tapped density measurement guide.
Check unexpected readings before interpreting them
If tapped density is lower than untapped density, or final tapped volume exceeds initial volume, verify the records, units, sample identity, possible material loss and test procedure. Do not turn an inconsistent reading into an apparently favorable flow classification. If a reading cannot be reproduced, retain the deviation and repeat according to the approved laboratory method.
Why Identical Hausner Ratios Can Produce Different Fill Weights
Consider two hypothetical blends with the same HR and CI but different absolute densities. Their relative consolidation is identical; their bulk masses per unit volume are different.
| Parámetro | Blend A | Blend B |
|---|---|---|
| Untapped bulk density | 0.500 g/mL | 0.640 g/mL |
| Tapped bulk density | 0.625 g/mL | 0.800 g/mL |
| Hausner ratio | 1.250 | 1.250 |
| Carr index | 20.00% | 20.00% |
| Illustrative mass in 1 mL at untapped density | 500 mg | 640 mg |
At the stated untapped densities, Blend B contains 28% more mass in the same illustrative 1 mL volume. This arithmetic comparison is not a prediction of tablet weight. Real die filling can change with aeration, dosing, feeder operation and consolidation, so the actual difference must be measured on the press.
This is why a useful pre-tableting report retains bulk and tapped density as numerical values rather than recording only the two flow indices. Identical HR and CI readings do not establish identical filling volume, compression behavior or machine suitability.
How Hausner Ratio Relates to Rotary Tablet Press Performance
A rotary tablet press fills a defined die cavity, adjusts the quantity of powder, compacts the bed and ejects the tablet. A powder’s flow behavior and packing state can influence the mass available during filling. Consequently, an unexpected laboratory index may help engineers decide which process conditions to investigate.
Feeder and turret conditions
Schomberg, Kwade and Finke studied five formulations on a rotary press and examined how material properties and operating settings influenced filling. Their work found that short filling times at higher turret speeds could cause incomplete filling; paddle speed and dosing conditions could also change the density of the powder in the die. [3] These findings support evaluating filling at the intended turret and feeder settings rather than selecting a speed from the flow index alone.
Investigate the source of tablet weight variation
| Observación | Useful next check |
|---|---|
| Indices change between batches | Verify comparable sample preparation, density readings and material history |
| Same HR but tablet mean weight changes | Compare absolute densities, dosing depth and actual feed behavior |
| Tablet weight spread increases as speed rises | Review available filling time, feeder behavior and recorded settings |
| Stable weight but different breaking force | Investigate formulation and compression conditions separately |
| Powder bridges or discharges intermittently | Use an appropriate discharge or shear test and examine hopper conditions |
Changing several settings at once makes it difficult to identify the cause of an observed change. Use an agreed trial plan, record relevant settings, and examine individual tablet weights and variability. Additional checks may include tablet hardness testing and the observations described in our tablet defects guide.
Hausner Ratio vs. Angle of Repose and Other Flow Tests
Choose a test based on the problem to be investigated. The Hausner ratio and Carr index describe the relative change in bulk packing under tapping. The angle of repose describes the slope formed by powder under a selected pile-forming method. Discharge and shear tests characterize other aspects of movement and consolidation.
| Measurement | Result | Practical use |
|---|---|---|
| Bulk and tapped density | Mass per apparent bulk volume | Retain absolute values for packing comparisons |
| Hausner ratio / Carr index | Relative consolidation | Screen and compare paired density data |
| Angle of repose | Slope of powder pile (degrees) | Compare relative resistance to movement under the test method |
| Flow through an orifice | Discharge behavior or mass flow rate | Investigate flow through an opening |
| Shear testing | Stress-dependent flow behavior | Investigate cohesive material behavior under defined loads |
In a 2024 comparative study of 21 powders, Tharanon and colleagues found that the pharmacopoeial methods differed in how well they distinguished materials across the flowability range. They recommended using several techniques when feasible. [4] When the indices and another flow measurement disagree, check the methods and the handling problem they describe. It may be reasonable for a powder to exhibit limited volume change during tapping yet behave differently through a particular opening. Read our angle of repose test guide for cone geometry, measurement procedure and its limitations.
Powder Flow Records Before a Tablet Press Trial
Prepare a concise material record before discussing equipment. Include representative sample identity, formulation characteristics, both measured densities, calculated HR and CI, the measurement method and repeat results. If the material cannot be disclosed in detail, provide the agreed non-confidential characteristics needed for engineering review.
For the trial itself, agree on target tablet weight and dimensions, production objectives, the proposed machine and tooling configuration, operating records and quality evaluation criteria. Capture any observed feeding interruption or refill event alongside the collected tablet samples.
Connecting the material record to equipment
El GRAND ZP31D Rotary Tablet Press is one example of downstream equipment in the GRAND tablet press range. The laboratory tests discussed here use separate measurement apparatus. Suitability for the intended formulation must be determined from the equipment configuration, tablet specifications and relevant trial evidence, not solely from an HR or CI number.

Explore the GRAND rotary tablet press range when comparing machines and preparing a technical brief for an equipment trial.
Preguntas frecuentes
What is a good Hausner ratio for powder flow?
The referenced relative flow ranking describes 1.12–1.18 as good, and 1.00–1.11 as excellent. These ranges are descriptive. Actual acceptance criteria depend on the formulation and the validated manufacturing process.
What is the formula for Carr index?
CI (%) = 100 × (tapped density − untapped bulk density) / tapped density. For volumes measured on the same unchanged sample, use 100 × (initial volume − final tapped volume) / initial volume.
What is the difference between Hausner ratio and Carr index?
Hausner ratio is dimensionless; Carr index is a percentage. Both are computed from the same paired density or volume readings. Their mathematical relationship is CI = 100 × (1 − 1/HR).
Can the Hausner ratio be less than 1?
Under the intended test conditions the tapped volume normally does not exceed the starting volume, so HR should be at least 1. Review the sample and method if a lower value is obtained.
Does a Carr index of 20% mean the powder can be compressed into tablets?
No. In the referenced relative ranking, 20% is within the fair-flow band. Tabletability, die filling and finished tablet quality require separate assessment.
Does the Hausner ratio predict tablet hardness?
It cannot determine tablet breaking force. Formulation, compaction conditions, tooling and the resulting tablet structure also matter.
Should I measure both Hausner ratio and Carr index?
Both can be calculated from one correctly paired set of density readings. Report the original readings and consider independent measurements when the manufacturing question requires more information.
Conclusión
The Hausner ratio and Carr index provide two ways to express relative powder consolidation from the same bulk and tapped measurements. Calculate them using consistent units and a documented method. Retain the original densities, compare replicate measurements and investigate discrepancies with complementary tests. For a rotary tablet press project, the most useful evidence connects this laboratory record to actual feeding behavior, die filling and tablet quality under documented trial conditions.
Referencias técnicas
- United States Pharmacopeia. General Chapter 〈1174〉 Powder Flow. Stage 4 harmonized revision, official May 1, 2024. DOI: 10.31003/USPNF_M99885_02_01.
- Akseli I, Hilden J, Katz JM, et al. Reproducibility of the Measurement of Bulk/Tapped Density of Pharmaceutical Powders Between Pharmaceutical Laboratories. Journal of Pharmaceutical Sciences. 2019;108(3):1081–1084. DOI: 10.1016/j.xphs.2018.10.009.
- Schomberg AK, Kwade A, Finke JH. The Challenge of Die Filling in Rotary Presses—A Systematic Study of Material Properties and Process Parameters. Pharmaceutics. 2020;12(3):248. DOI: 10.3390/pharmaceutics12030248.
- Tharanon W, Guo Y, Peerapattana J, Sun CC. A systematic comparison of four pharmacopoeial methods for measuring powder flowability. International Journal of Pharmaceutics. 2024;661:124454. DOI: 10.1016/j.ijpharm.2024.124454.